Paints and inks often exhibit different shear rates from manufacturing to application. As typical non-Newtonian fluids, paints and inks show varying viscosity characteristics at different shear rates. Generally, paints are at a low shear rate when stored, transported, leveled, or sagged. They experience medium shear rates when pumped, dipped, or mixed at low speeds. During high-speed dispersion, rolling, spraying, and brush coating, they are subjected to high shear rates, generally ranging from 9000 s⁻¹ to 12000 s⁻¹. Therefore, understanding the rheological characteristics of paints or inks at such high shear rates requires the use of a cone-and-plate viscometer for measurement.
The BGD 182 Cone-and-plate Viscometer uses a specially angled conical spindle, which generates a very high shear rate on the sample under the high-speed drive of a stepper motor. It conforms to ISO 2884-1: "Paints and varnishes — Determination of viscosity using rotary viscometers — Part 1: Cone-and-plate viscometer operated at a high rate of shear."
Standards
ISO 2884-1 (Paints and varnishes - Determination of viscosity using rotary viscometers - Part 1: Cone-and-plate viscometer operated at a high rate of shear)
Technical Specification
7-inch touch screen: Menu operation with rich display content (measurement values, spindle number, speed, shear rate, etc.), simple and convenient to operate
Durable metal shell: Handle for lifting, precise and fast positioning, high reliability
Carefully designed electrical part: Ensures high reliability, precision, stability, and ease of use
ARM chip processor: Faster data processing speed
High precision linear calibration: Ensures higher measurement accuracy through multiple point interpolation
Stepless speed adjustment: Allows operators to choose different shear rates based on the sample's actual application conditions
Continuous viscosity measurement: Many viscosity units can be freely interchanged
Automatic measurement alarm: Alerts when the measurement exceeds the range
High-speed data transmission interface: Ensures fast and stable communication between the instrument and the computer
Data storage and export: External USB function
Built-in high-precision PT100 temperature sensor: Provides high temperature control accuracy and stability
Temperature calibration function: Ensures precise and reliable temperature control
Simple calibration function: Users can quickly calibrate the instrument with one bottle of standard oil using the built-in calibration menu
Data collection and analysis software: Option to thoroughly analyze the sample's rheological features